510(k) vs De Novo vs PMA: Which FDA Pathway Fits Your AI Medical Device

10 Jul, 2026

FDA pathways for AI medical devices: 510(k), De Novo, and PMA

By the end of 2025, the FDA had authorized 1,451 AI and machine learning-enabled medical devices. Roughly 95 percent of them cleared through one pathway: 510(k). De Novo accounted for most of the rest. PMA accounts for a small fraction, in line with how few AI devices carry Class III risk.

The split comes down to risk class and whether a predicate device exists. Once you understand those two factors, the pathway decision gets a lot simpler.

This post walks through what qualifies as an AI-enabled SaMD or SiMD, how the three pathways differ, and how to figure out which one applies to your device.

What counts as an AI-enabled SaMD or SiMD

An AI-enabled medical device uses artificial intelligence or machine learning for a medical purpose. That purpose may include detection, diagnosis, triage, monitoring, prediction, measurement, or clinical decision support.

The AI function may work as standalone software. It may also sit inside a physical device.

AI-enabled SaMD means the AI function works as Software as a Medical Device. FDA follows the IMDRF definition of SaMD. It refers to software used for one or more medical purposes without being part of a hardware medical device.

Examples include:

  • AI software that detects diabetic retinopathy from retinal images
  • AI software that flags findings in CT or MRI scans
  • AI software that supports stroke triage
  • AI software that analyzes ECG signals
  • AI software that supports clinical risk assessment

AI-enabled SiMD means the AI function sits inside or supports a physical medical device.

Examples include:

  • AI reconstruction software in an imaging system
  • AI guidance software in a surgical system
  • AI monitoring software in a wearable device
  • AI automation inside diagnostic equipment

Not every AI healthcare tool is a medical device. The FDA excludes certain categories, like administrative software, general wellness apps, and basic clinical decision support that lets a clinician independently review the recommendation.

If the software is intended to diagnose, treat, or otherwise affect a clinical decision, it falls under FDA oversight, and the pathway question applies. If it does not, the FDA's device pathways do not come into play at all.

From here, pathway choice starts with risk and predicate status.

The three pathways

510(k) clears a device by showing it is substantially equivalent to a device already on the market, called a predicate. You are not proving the device is safe and effective from scratch. You are proving it works the same way as something the FDA has already reviewed.

De Novo applies to novel devices with no predicate, where the risk is low to moderate. Instead of comparing to an existing device, you ask the FDA to create a new classification. Once granted, that classification can become a predicate for future 510(k) submissions, including from other companies.

PMA, or Premarket Approval, applies to Class III devices. These carry the highest risk, and PMA requires the most clinical evidence of the three pathways. It is closer to a full safety and effectiveness review than a comparison exercise.

Underneath all three sits the FDA's risk classification system:

  • Class I: low risk, minimal controls
  • Class II: moderate risk, most AI-enabled devices land here
  • Class III: high risk, or the device supports or sustains life

The pathway follows the risk class. Class I and II devices with a predicate go through 510(k). Novel Class I or II devices go through De Novo. Class III devices go through PMA.

How to tell which one applies to your device

Start with the intended use.
Then check predicate status.
Then assess risk.

A predicate exists for the same intended use, and the device is Class I or II → 510(k). Class III devices go through PMA regardless of whether a predicate exists. No predicate exists, and the risk is low to moderate → De Novo.

The claim matters more than the model itself. The same AI model can land in different pathways depending on what it claims to do. A tool that ranks images for a radiologist to review carries a different risk profile than one that diagnoses disease on its own, even if the underlying model is identical.

LumineticsCore, formerly known as IDx-DR, shows this well. It was the first autonomous AI diagnostic system the FDA authorized, cleared in 2018 to detect diabetic retinopathy without a clinician reviewing the image. No predicate existed for an AI system making an unsupervised diagnostic call. It went through De Novo. That clearance then became the predicate other diabetic retinopathy tools later used to clear through 510(k).

Device situation

Likely pathway

Predicate exists for the same intended use

510(k)

No predicate exists, but risk is low to moderate

De Novo

Device is Class III or high risk

PMA

Device supports or sustains life

PMA

Device has planned AI updates after authorization

PCCP may apply

What each pathway costs in time

Timelines vary by device complexity and submission quality, but the pattern holds across pathways.

510(k)

The FDA's MDUFA V goal targets a 90 FDA-day review for most submissions, with a total time to decision goal of 112 calendar days. In practice, the median review time for AI and ML-enabled devices authorized in 2024 was about 151 days.

De Novo

FDA's MDUFA V goal targets a 150 FDA-day decision for the majority of requests by FY2026 and FY2027. Actual median review time for AI and ML devices in 2024 ran closer to 372 days, reflecting the added work of building a case with no predicate to lean on.

PMA

The shared outcome Total Time to Decision goal is 285 calendar days for original PMA submissions received in FY2025 through FY2027. This figure covers both FDA review time and the time sponsors spend responding to information requests, so total calendar time is usually longer than the number suggests.

510(k) is fastest because it leans on existing evidence. De Novo and PMA both require you to build the evidence case yourself, and that shows up directly in the timeline.

Where training data enters the decision

The pathway changes what your evidence needs to prove.

For 510(k), the data must support substantial equivalence.

  • You need to show that the AI device performs safely and effectively for the same intended use as the predicate.

For De Novo, the data must support a new classification.

  • You need to prove that the device is safe and effective without relying on a predicate.

For PMA, the data must support direct approval.

  • You need a strong safety and effectiveness case.

This is where many AI teams underestimate the work. FDA will want to understand the data behind the model.

That includes:

  • Where the data came from
  • Who collected it
  • How it matches the intended population
  • How ground truth was created
  • Who annotated or reviewed the data
  • How datasets were split
  • How the test set stayed independent
  • How the model performs across subgroups
  • How the model handles edge cases
  • What the labeling says about limitations

The pathway changes the argument. It does not reduce the need for data quality.

Common mistakes that slow this decision down

Assuming a predicate exists

A similar device is not always a valid predicate. The intended use, user, input data, output, and workflow must align.

Treating AI as the pathway driver

AI does not decide the pathway. Risk and predicate status do. An AI device can go through 510(k), De Novo, or PMA.

Ruling out PMA too early

PMA is uncommon for AI-enabled devices, but a Class III device still needs Class III evidence: a prospective clinical trial, a full safety and effectiveness case, and often an FDA advisory panel review. 

Assuming your device won't need this, because most AI devices don't, is how teams under-resource a PMA submission that was always going to require it.

Making a novel claim with 510(k) evidence

A new claim may need more than equivalence. If the intended use moves beyond the predicate, the evidence burden changes.

Underestimating training data expectations

Training data must match the claim. FDA needs to see how the data supports the intended use.

Getting the pathway right

AI devices rarely fail because of a bad pathway choice. They fail because the evidence behind that choice was thin.

Pick the pathway based on risk and predicate status, not on speed. A faster pathway you do not qualify for will cost you more time than a slower one done right the first time.

Once the pathway is set, the real work starts. You need to prove your training data supports the claim you are making, whether that claim is equivalence to a predicate or a case built from the ground up.

The AI SaMD Data Playbook walks through what regulatory-ready training data actually looks like: how to build the dataset, evaluate for bias and subgroup performance, document everything for traceability, and handle updates under a PCCP.

FAQ

Is FDA clearance the same as FDA approval for AI medical devices?

No. 510(k) and De Novo result in FDA clearance or authorization. PMA results in FDA approval. The terms are not interchangeable, and regulatory language in marketing materials should reflect which one applies.

Does using AI automatically make a medical device higher risk?

No. Risk classification depends on intended use and clinical consequence of error, not on whether AI is involved. Many AI enabled devices are Class II, the same class as many non-AI devices with similar functions.

Can an AI medical device change its algorithm after FDA clearance?

Only within the bounds of an approved PCCP, or Predetermined Change Control Plan. Changes outside that scope, such as a new indication or a materially different clinical claim, need a new submission.

How many AI and machine learning medical devices has the FDA authorized?

By the end of 2025, the FDA had authorized 1,451 AI and ML enabled medical devices, with 295 authorized in 2025 alone.

Does an AI device need a new FDA submission every time the model is retrained?

Not necessarily. Retraining within the bounds of an authorized PCCP does not require a new submission. Retraining outside that scope, or without a PCCP in place, does.